Determination of the radiochemical and chemical purity of the radiopharmaceutical drug prostate-specific membrane antigen labeled with lutetium-177 (177Lu-PSMA-617)
DOI:
https://doi.org/10.52304/.v28i3.668Keywords:
radiopharmaceuticals, PSMA-617, prostate cancer, radiochemical purity, paper chromatography, atomic emission spectrometryAbstract
The work is devoted to the development of techniques for determining the radiochemical purity (the content of the connected target radionuclide) and chemical purity (the content of nonradioactive impurity elements Cu, Fe, Pb, Zn, Yb, etc. of the radiopharmaceuticals 177LuPSMA-617, produced in the State Enterprise "Radiopreparat". At the development of the techniques, the methods of paper chromatography and atomic emission spectrometry with inductively coupled plasma (ICP-AES) are used. The paper chromatography method made it possible to determine the radiochemical purity of the preparation up to 99.0%, and ICP-AES determined the limited impurities of Cu, Fe, Pb, Zn, Yb up to 0.35 µg/GBq. The characteristics of the developed methods fully meet the regulatory requirements established for radiopharmaceuticals.